7. Cooper protocol,
e.g. for ECG stress test
N
Start at 5.3 km/h and 0% elevation
N
after 1 minute elevation increases to 2 %
N
after another minute, the elevation is increased by 1% every minute
N
when elevation is 25 % elevation stays constant and the speed is
increased by 0.32 km/h every minute
STOP must be activated manually by the medical doctor when the subject is
completely exhausted.
8. Ellestad A protocol,
e.g. for ECG stress test
Step
Duration (min)
Speed (km/h)
Elevation (%)
1
3:00
2.7
10.0
2
4.8
3
6.4
4
8.0
9. Ellestad B protocol,
e.g. for ECG stress test
Step
Duration (min)
Speed (km/h)
Elevation (%)
1
3:00
2.7
10.0
2
4.8
10.0
3
6.4
10.0
4
8.0
10.0
5
8.0
15.0
6
9.6
15.0
10. Ramp profile
(not available for every model)
Ramp profile with 2 parameters:
N
Target speed standard: 10.0 km/h; adjustable from 0 to maximum
speed of the treadmill.
N
Time for reaching target speed in seconds: standard: 10 seconds;
adjustable from 0 to 99 seconds