safety, installation and commissioning
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 113 of 216
[05]
Perform an electric safety test/measurement for
N
protective earth resistance
N
isolation resistance and
N
leakage current
Record the test results in a special protocol [cos11690xx] and
mark them as “first measured values“. One copy of this protocol
remains with the owner’s manual and the original is sent to
h/p/cosmos.
The potential equalization cable has to be connected AFTER these
measurements.
[06]
Lubricate the running machine with 30 ml special silicone oil, which
is part of the “service box”.
[07]
Check the tension of the running belt (no slip).
[08]
Adjust the running belt properly before instructing the customer.
The allen key must be pulled out of the screw
immediately after usage due to risk of injury!
Make sure that nobody exposes body parts, hair, clothes, ties,
towels or other parts to any dangerous capture areas/gaps at
moving parts such as the rollers or elevation system.
[09]
Test the function of the heart rate transmission with the help of a
simulator or a POLAR sender. Search for interferences and
possible reasons if applicable.
accomplished
accomplished
accomplished
accomplished
accomplished