optional settings: administrator options
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 151 of 216
OP33
Distance between rungs of the
ladder ergometer
(option only available for ladder
ergometers, not for treadmills)
For correct calculation of the number of rungs / steps, the distance
between two rungs can be changed and confirmed in this option.
Standard: 254 for 25.4 cm
OP34
Manual speed calibration
(counted increments from the speed
sensor for 10 m belt movement)
Note.
Normally the speed calibration
values are set by selecting the
correct device type (administrator
option number 3).
This value is permitted to be changed only if the final drive ratio
changes.
If you change the speed calibration value, please also pay attention to
the correct settings in OP5 (firmware speed limit), OP11 (speed
measurement; must be switched on) and OP48 (frequency inverter
speed limit) of the administrator mode. To change the speed
calibration value, please use the + and – buttons. For quick
movements between standard values, please use the UP and DOWN
buttons.
If the speed measurement is switched off in OP11, OP34 will only
report internal program absolute values. Otherwise, you can see the
stored values in the list of running machines types in the appendix.
OP35
Oil interval
The distance interval for indication of required "OIL" service can be
adjusted by means of OP 35.
Standard = 1000 km for treadmills, 100 km for ladder ergometers.
OIL intervals are indicated with error code "E01" "OIL" every 1000
km.
§
indicates oil interval in km (standard = 1.000)
km is flashing, adjustment range: OFF, 100 ... 5.000 km
Do not change the interval without prior communication with
h/p/cosmos. For the models h/p/cosmos venus and saturn, JAEGER
CareFusion LE 580 CE and larger as well as COSMED T200 and
larger, OP35 has the standard setting OFF because the machines are
equipped with an automatic oil pump and tank.
OP37
Interval for service message TSC
(time-dependent)
for technical safety control
E.02 “HELP"
On sports devices and/or active medical devices with power supply
connection used for either commercial, industrial, institutional and
medical applications, regular technical safety controls have to be
performed by a trained and authorised service technician. Appropriate
protocols and inspection instructions are available at h/p/cosmos
service. An error message "E 02" "HELP" for the TSC-service interval
will appear in the display after expiration of a time period of 12
months.
The remaining time period (in months) till the next TSC / maintenance
/ inspection can be viewed by calling up OP37.
Reset of the interval and the error report „E02“ is possible via option
OP47.
The TSC report interval can be changed via OP37:
Select OP37 and change with
a
or
d
.
Confirm with
c
.
OP37 can also be used for the adjustment of the maintenance
intervals in months if the mileage is small.
§
indicates: TSC-Interval in months (standard = 12 months)
Adjustment range: OFF, 0 to 48 months