PSU Series Programming Manual
60
2.
Press the Function key to enter the
Normal configuration settings.
Page 22
Set the following LAN settings:
F-36 = 1
Turn LAN on
F-37 = 1
Enable DHCP
F-59 = 1
Turn the web server on
F-60 = 0 or 1
Set to 0 to disable web password,
set to 1 to enable web password
F-61 = 0000
~9999
Set the web password
3.
The LAN indicator will turn on when a
network cable is plugged in.
LAN indicator
A
V
VOLTAGE
CURRENT
VSR
LAN
RMT
ERR
DLY
ALM
ISR
M 1
M 2
M 3
RUN
C V
C C
Note
It may be necessary to cycle the power or refresh the
web browser to connect to a network.
Web Server Remote Control Function Check
Functionality
check
Enter the IP address of the power supply in a web
browser after the instrument has been configured
as a web server (page 59).
The web server allows you to monitor the function
settings of the PSU.
You can check the IP address by checking F-39 to F-42.
F-39 = AAA
IP Address part 1 of 4
F-40 = BBB
IP Address part 2 of 4
F-41 = CCC
IP Address part 3 of 4
F-42 = DDD
IP Address part 4 of 4