REMOTE CONTROL
57
Selects the unit with address 8 and
returns its identity string.
Note
ADR 11\r
Typing
OK\r
Return
IDN? \r
Typing
GW-INSTEK,PSU6-200,,T0.01.12345678\r Return
Selects the unit with address 11 and
returns its identity string.
Note
Note
TDK commands do not use LF (line feed) codes to
terminate commands. See the TDK GENESYS user
manual for further information.
Multi-Drop mode
When using the Multi-Drop mode, the entire SCPI
command list developed for the PSU can be used.
Each unit can be individually controlled after a
slave unit has been selected. For this function
check, we will assume that the master unit is
assigned to address 0, while a slave is assigned
address 5.
Run this query command via the terminal
application after the instruments have been
configured for multi-unit control with Multi-Drop
mode. See page 51.
SCPI commands
Command or response
Status
INST:SEL 0
Typing
*IDN?
Typing
GW-INSTEK,PSU150-10,,T0.01.12345678 Return
Selects the unit with address 0 and
returns its identity string.
Note
INST:SEL 5
Typing
*IDN?
Typing