GETTING
STARTED
17
Power up
Panel operation
1.
Set the rear panel
Voltage selector to the
correct position
according to the AC
mains voltage.
115V
115
V
230V
230V
2.
Connect the power
cord to the socket.
3.
Turn On the power
switch. The display
becomes active in 2~3
seconds.
DC ,
20 1
M Hz
Preci
sion
LCR
Met
er
4.
Use the contrast knob on the rear panel to
adjust the LCD display brightness.
Counter-
clockwise:
Dark
Clockwise:
Light
MEASUREMENT MODE
0.01234
0.2179
nS
D
*7
C
L X B Z Y
Q
D R G
Parallel
Hide Scale
Abs
%
Save Nom
HI
53.388fF
Hi
Lo
+1.00%
–1.00%
50mVac
1.5000kHz
Range
Speed Med
Auto
Vm:549.8mV
Im:724.9pA
–
MEASUREMENT MODE
0.01234
0.2179
nS
D
*7
C
L X B Z Y
Q
D R G
Parallel
Hide Scale
Abs
%
Save Nom
HI
53.388fF
Hi
Lo
+1.00%
–1.00%
50mVac
1.5000kHz
Range
Speed Med
Auto
Vm:549.8mV
Im:724.9pA
–
LCR-8101
User
Manual
18
Select AC mains frequency (50/60Hz)
Background
Although LCR-8101 works under both 50 and
60Hz power frequencies, we recommend selecting
the frequency that matches the local setting to get
the best measurement precision, especially at
lower frequencies (< 100Hz).
Panel operation
1.
Press the Menu key, then F5
(System). The system menu appears.
Menu
F 5
Precision LCR Meter LCR-8101
5
:
Frequency
Software version 2.00 Jan 20 2007
RS-232
GPIB
Graph mode
Line frequency
Beep
GPIB address
:
:
OFF
50Hz
1MHz
2.
Press the Up/Down key and move
the cursor to Line frequency.
:
Line frequency
50Hz
3.
If necessary, press the Left/Right
key to select the frequency, from 50
to 60Hz or from 60Hz to 50Hz.
:
Line frequency
60Hz