BASIC
MEASUREMENT
47
Select frequency
step resolution
For frequency increase/decrease using Up/Down
keys, fine and coarse step settings are available.
Fine
1
st
digit: 1, 2, 3, 4, 5, 6...
Coarse
2
nd
digit: 10, 12, 15, 20, 25, 30, 40, 50, 60, 80
1.
Press the Code key.
Code
2.
Enter the system code using the numerical
keys, then press the Enter key. A confirmation
message appears on the display.
Fine:
10
Enter
Freq fine steps
Coarse:
11
Enter
Freq coarse steps
LCR-8101
User
Manual
48
Set measurement voltage
Background
The measurement voltage, together with the
measurement frequency, defines the electrical
condition of each measurement item. Make sure
the appropriate voltage is selected, according to
the component characteristics.
Frequency setting
1.
Press the Left/Right key repeatedly
to move the cursor to Voltage.
195.00
2.00 Vac
kHz
2.
Enter the voltage using the numerical keys.
Range
10mV ~ 2V
100mV
m
V/A
Enter
1V
V/A
Enter
Backspace
All clear
Clear
Increase
Decrease
When the entered value does not fit in the
range, LCR-8101 automatically selects the
nearest value.
Nearest Available
When the wrong unit (such as
Ω
) is entered,
the value is cancelled.
Unit Mismatched