PASS-FAIL
MODE
59
Result > Hi
HI
Result < Lo
LO
Lo < Result < Hi
(Pass)
PASS
Run in Percentage mode
1.
Press F5 to select Percentage
measurement.
F 5
Abs
%
2.
Use the Left/Right key to move the
cursor to Hi/Lo value for editing.
Hi
Lo
+1.00%
–1.00%
100.00mD
3.
Use the numerical keys
and unit keys to enter the
value. The editing value
appears at the bottom left
corner of the display.
Range
Speed Med
Auto
Lo Lim: 1.5_
–2.50%
Enter
1.5kH
Enter
H
k
Backspace
All clear
Clear
Increase
Decrease
LCR-8101
User
Manual
60
Hi and Lo values are automatically swapped if
necessary
Hi and Lo Swapped
4.
The display updates the Hi/Lo result
immediately. The result is pass if the bar stays
in the central area. The buzzer sounds
accordingly.
Result > Hi
HI
Result < Lo
LO
Lo < Result < Hi
(Pass)
PASS
Use display value as nominal
Panel operation
(For Percentage only) The displayed
value can be used as the nominal value.
Press F6 (Save Nom) to copy the
display value to the nominal value.
F 6
1.2345
0.2179
mF
D
LO
–
Hi
Lo
+1.00%
–1.00%
100.00mF
1.2345
0.2179
mF
D
PASS
–
Hi
Lo
+1.00%
–1.00%
1.2345mF