PASS-FAIL
MODE
61
Multi-Step Test Configuration
Overview
Background
The multi-step function can configure and run
multiple measurement steps. Maximum 64
programs, 30 steps each, can be programmed and
stored in the instrument.
Limit type
Only the absolute limit
testing is available. For
percentage limit test, use
the single mode (page53).
Low
Limit
Hi
Limit
Pass
C
S
Series capacitance
X
Reactance
Test item
C
P
Parallel capacitance
B
Susceptance
L
S
Series inductance
Z
Impedance
L
P
Parallel inductance
Y
Admittance
R
S
Series resistance
R
DC
DC Resistance
R
P
Parallel resistance
θ
Angle
For detailed description of each item, see page31.
Parameters
No. of step
Maximum 30 for each program
No. of program
Maximum 64
Drive Voltage
10mV ~ 2V, 1mV step
Frequency
20Hz ~ 1MHz
Bias
Reserved item: internal use only
Speed
Max, Fast, Med, Slow
Hi / Lo Limit
Follows the measurement range
Delay
0 ~ 9999ms, 1ms step
Single trigger
Program runs when the Trigger
key or F1 (Start) is pressed.
Automatic
trigger
Program runs when LCR-8101
detects the DUT.
LCR-8101
User
Manual
62
Configure beep setting
Background
The beep sounds when the pass/fail test result
matches the setting: failed or passed.
Panel operation
1.
Press the Menu key, then F5
(System). The system configuration
appears.
Menu
F 5
Precision LCR Meter LCR-8101
5
:
Frequency
Software version 2.00 Jan 20 2007
RS-232
GPIB
Graph mode
Line frequency
Beep
GPIB address
:
:
OFF
50Hz
1MHz
2.
Press the Up/Down key and move
the cursor to Beep.
Beep
:
OFF
3.
Press the Left/Right key to select the
beep setting, Off, Pass, or Fail.
Off
Beep is turned Off
Pass
Beeps when the test result is pass
Fail
Beeps when the test result is fail