REMOTE
CONTROL
97
Select GPIB
address
1.
Press the Menu key and F5 (System).
The system configuration appears.
Menu
F 5
Precision LCR Meter LCR-8101
5
:
Frequency
Software version 2.00 Jan 20 2007
RS-232
GPIB
Graph mode
Line frequency
Beep
GPIB address
:
:
OFF
50Hz
1MHz
2.
Press the Up/Down key and move
the cursor to GPIB.
5
GPIB address
:
3.
Use the numerical keys to enter the GPIB
address, 1 ~ 30.
30
GPIB address
:
Address
5
Enter
LCR-8101
User
Manual
98
Command Syntax
Compatible
standard
•
IEEE488.2, 1992 (fully compatible)
•
SCPI, 1994 (partially compatible)
Command format
trig:del:mod <NR1>LF
1
2
3
4
1: command header
2: single space
3: parameter
4: message terminator
Parameter Type
Description Example
<Boolean>
boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal numbers 0.1, 3.14, 8.5
<NR3>
floating point
4.5e-1, 8.25e+1
<disc>
discrete data
on, off, max
Terminates a command line. Note that LCR-8101
accepts only LF (line feed) as the terminator.
Message
terminator
LF
line feed code
Note
•
Commands introduced here are described in
abbreviated style (same functionality)
•
Commands are non-case sensitive.