
GRAPH
MODE
81
If a wrong unit is entered, the value is cancelled.
Unit Mismatched
If a value outside of the range is entered, the
closest available value is automatically selected.
Nearest Available
If the entered start Voltage level is higher than
the stop Voltage, the two values are swapped.
Hi and Lo Swapped
Set stop
Frequency
4.
Repeat the above step for the stop Frequency.
Stop:
1.00 V
Range
20Hz ~ 1MHz (stop Frequency must
be higher than the start Frequency)
Set measurement
Voltage
5.
Use the Up/Down key to move the
cursor to Voltage setting (Level).
Level:
1.00 V
Use the numerical keys to enter the
measurement Voltage.
Range
10mV ~ 2V
100mV
m
V/A
Enter
1V
V/A
Enter
Backspace
All clear
Clear
Increase
Decrease
Select Lin/Log
scale
6.
Press F1 to select the horizontal
scale, Linear or Logarithmic.
Log(Hz)
Lin(Hz)
F 1
LCR-8101
User
Manual
82
Vertical Scale Setting
Set vertical axis ( Absolute)
Background
The Y (vertical) axis configuration is available for:
•
Manual or Automatic fit: Selects whether the
vertical range is manually set or automatically
adjusted as the graph is plotted.
•
Absolute or Percentage: Selects how the vertical
range is defined, as absolute values (minimum
and maximum) or percentage of the nominal
(center) value.
Panel operation
1.
Press F2 to select Abs.
Abs
%
F 2
2.
Press F3 to select Manual Fit.
Manual Fit
F 3
3.
The maximum (Hi) and minimum (Lo) level of
the vertical axis appears.
Step Size:
Cs Hi
Cs Lo
:
:
2 3 4
1
3.5626mF
5.8240mF
Set Hi level
4.
Press the Up/Down key to move the
cursor to Hi level.
Cs Hi:
5.8240mF
5.
Use the numerical keys to enter the Hi level.
Range
According to each measurement item
(see page31).
1.2m
Ω
m
Ω
Enter