Inside the DM24
A terminal window opens, presenting you with the command line of the digitiser:
If you are using a different emulator, open a window as normal with the appropriate
Baud rate. The communications settings should be 8-N-1 with no flow control.
Check that there is two-way communication with the digitiser by pressing Enter.
The digitiser should reply with
ok
on a new line.
Type
re-boot
to reinitialize the digitiser, and confirm with
y
. As it is restarting, the
digitiser will report its status over the terminal connection, followed by a
maintenance menu:
MPE ARM ANS ROM PowerForth v6.30
ARM Serial BootStrap v1.100, 11 August 2003
Copyright (c) 2002-3 GSL, EDSL & MicroProcessor Engineering Ltd.
Port 0 38400 baud Port 1 4800 baud Port 2 38400 baud
Guralp Systems Ltd - ARM-BOOT v3.0 mgs 13/12/12 (Build _02)
System Code versions loaded :-
Current 0105:0000 Guralp Systems Ltd - DM+FW v.107 mgs 13/09/16 (Build _26)
Backup 010E:0000 Guralp Systems Ltd - DM+FW v.107 mgs 13/09/16 (Build _26)
Previous 0117:0000 Guralp Systems Ltd - DM+FW v.106 mgs 11/11/14 (Build 57w)
DSP Code :
0103:0000 dsp1091.bin loaded 2016-09-29 Default
0104:0000 dsp1091.bin loaded 2016-09-29
Command keys:
C - set real time Clock (2016 9 29 14:37:06 )
I - view/upload InfoBlock
F - run the Forth monitor
S - update System program
O - select Other system program
B - update Boot program
D - update DSP code
T - Toggle default DSP code
Q - Quit maintenance system
5 seconds to auto-start:
If you do not press a key in the next five seconds, the DM24 will start up normally. If
you need more time to read the menu, press the space-bar.
104
Issue U - December, 2021