Command line interface
6.9.8 -AUTOCAL
Syntax:
-AUTOCAL
Cancels any automatic calibration setting in force. If automatic calibration is not
enabled, this command has no effect.
6.10 Actions
6.10.1 LOCK
Syntax:
LOCK
or
LOCK MONITOR
Locks the sensor mass(es) of the instrument connected to
SENSOR A
.
If you issue LOCK, the command returns immediately, allowing you to close the
terminal connection and follow the progress of the lock from the data streams.
If you issue
LOCK MONITOR
, the system displays the mass position for each
component, measuring one position each second (
i.e.
one reading every 4 seconds
for each component of a 4-component instrument). The process returns when
locking is complete.
The mass positions are measured to 16-bit accuracy, with full deflection
corresponding to values around ±32000 counts.
6.10.2 UNLOCK
Syntax:
UNLOCK
Unlocks the sensor mass(es) of the instrument connected to
SENSOR A
.
UNLOCK
returns immediately, whilst
UNLOCK MONITOR
monitors the progress of the
unlock by displaying mass positions, as for
LOCK MONITOR
(above).
6.10.3 CENTRE
Syntax:
CENTRE
or
CENTER MONITOR
Centres the sensor mass(es) of the instrument connected to
SENSOR A.
CENTRE
returns immediately, whilst
CENTRE MONITOR
monitors the progress of the
unlock by displaying mass positions, as for
LOCK MONITOR
(above).
When the masses are correctly centred, the mass positions should read less than
±1000 counts.