Inside the DM24
In the case where a single power supply is used to power all of the equipment, this is
not an issue and pin E should be connected directly to pin B on all units. Pin B
should then also be connected to the power supply's positive terminal.
The more common case, where each unit has it's own power supply, is shown below.
Power supply wiring is omitted for the sake of clarity but note that all the power
supply's ground terminals will be connected together via the links between pin A on
each unit. The power supplies must, therefore, have floating outputs or grounded
negative terminals.