30
Disassembling steps
a) Press the push-button and pull out the shank/file.
b) Sliding it in the direction of the socket of handpiece and then
removed with contra-angle. Otherwise it will be damaged.
c) When removing the protective silicon cover, pull it straight out
slowly.
d) When inserting and removing the contra-angle, turn thehandpiece
power off beforehand.
7.4.3 Cleaning
The cleaning should be performed no later than 24 hours after the
operation.
The cleaning can be divided into automated cleaning and manual
cleaning. Automated cleaning is preferred if conditions permit.
7.4.3.1Automated cleaning
•The cleaner is proved to be valid by CE certification in accordance
with EN ISO 15883.
▪There should be a flushing connector connected to the inner cavity
of the product.
▪The cleaning procedure is suitable for the product, and the irrigating
period is sufficient.
It is recommended to use a washer-disinfector in accordance with
EN ISO 15883. For the specific procedure, please refer to the automated
disinfection section in the next section "Disinfection".
Notes:
a) The cleaning agent does not have to be pure water.It can be
Summary of Contents for Endo Radar Plus