38
RF emissions
CISPR 11
Group 1
The model Endo Radar Plus
uses RF energy only for its
internal function. Therefore,
its RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR11
Class B
The model Endo Radar Plus
is suitable for used in all
establishments, including
domestic establishments and
those directly connected to
the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
lEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
lEC 61000-3-3
Complies
Technical Description Concerning Electromagnetic Immunity
Table 2: Guidance & Declaration - electromagnetic immunity
Guidance & Declaration — electromagnetic immunity
The model Endo Radar Plus is intended for use in the electromagnetic
environment specified below. The customer or the user of the model Endo Radar
Plus should assure that It is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
lEC 61000-4-2
±8kV contact
±2, ±4, ±8,
±15kV air
±8kV contact
±2, ±4, ±8, ±15kV
air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for Input/
output lines
±2kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
lEC 61000-4-5
±0.5, ±1kV line
to line
±0.5, ±1, ±2kV
line to earth
±0.5, ±1kV line to
line
±0.5, ±1, ±2kV
line to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.