7 Cleaning, Disinfection and Sterilization
7.1 Foreword
For hygiene and sanitary safety purposes, the the contra-angle, the lip
hook, the file clip, the protective silicon cover and the touch probe must
be cleaned, disinfected and sterilized before each usage to prevent any
contamination. This concerns the first use, as well as all subsequent uses.
7.2 General recommendations
7.2.1 Use only a disinfecting solution which is approved for its
efficacy (VAH/DGHM-listing, CE marking, FDA and Health Canada
approval) and in accordance with the DFU of the disinfecting solution
manufacturer.
7.2.3 Do not place the contra-angle in a disinfectant solution or in an
ultrasonic bath.
Do not use chloride detergent materials.
7.2.4 Do not use bleach or chloride disinfectant materials.
7.2.5 For your own safety, please wear personal protective equipment
(gloves, glasses, mask).
7.2.6 The user is responsible for the sterility of the product for the
first cycle and each further usage as well as for the usage of damaged or
dirty instruments where applicable after sterility.
7.2.7 The water quality has to be convenient to the local regulations
especially for the last rinsing step or with a washer-disinfector.
7.2.8 To sterilize the endodontic files, refer to the manufacturer's
instructions for use.
7.2.9 The contra-angle needs to be lubricated after cleaning and
disinfection, but before sterilization.
7.3 Cleaning and disinfection steps for the motor handpiece, the
AC adapter and the base.
Before and After each use, all the objects that were in contact with
infectious agents should be cleaned using towels impregnated with
a disinfecting and detergent solution (a bactericidal, fungicidal and
aldehyde free solution) approved by VAH/DGHM-listing, CE marking,
FDA and Health Canada.
Warning: Do not sterilize the motor handpiece, the AC adapter
and the base.