29
sufficiently validated equipment and product-specific procedures are
used for cleaning/disinfection and sterilization, and that the validated
parameters are adhered to during every cycle.
Please also observe the applicable legal requirements in your country
as well as the hygiene regulations of the hospital or clinic, especially
with regard to the additional requirements for the inactivation of prions.
7.4.1.2Post-operative treatment
The post-operative treatment must be carried out immediately, no
later than 30 minutes after the completion of the operation. The steps are
as follows:
1. Remove the products from the base, and rinse away the dirt on the
surface of handpiece with pure water (or distilled water/deionized water);
2. Dry the products with a clean, soft cloth and place it in a clean tray.
Notes:
a) The water used here must be pure water, distilled water or
deionized water.
7.4.2 Preparation before cleaning
Steps:
Tools: tray, soft brush, clean and dry soft cloth.
1. Remove the shanks/files.
2. Remove the file clip, isolation sleeve,Contra-angle and connecting
wire from the handpiece in sequence, and then put them into a clean tray;
3. Use a clean soft brush to carefully brush lip hook, file
clip,protective silicon cover,touch probe, head and back cover of the
contra-angle until the dirt on surface is not visible. Then use soft cloth to
dry the products and put them into a clean tray. The cleaning agent can
be pure water, distilled water or deionized water.