39
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
(>95% dip in
UT.)
for 0.5 cycle
<5 % UT
(>95% dip in
UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in
UT)
for 250 cycles
<5 % UT
(>95% dip in UT.)
for 0.5 cycle
<5 % UT
(>95% dip in UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in UT)
for 250 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
models Endo Radar
Plus requires continued
operation during power
mains interruptions, it
is recommended that
the models Endo Radar
Plus be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
lEC 61000-4-8
30A/m
30A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Table 3: Guidance & Declaration - electromagnetic immunity concerning
Conducted RF & Radiated RF
Guidance & Declaration - Electromagnetic immunity
The model Endo Radar Plus is intended for use in.the electromagnetic
environment specified below. The customer or the user of the models Endo
Radar Plus should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance