28
Tools: Nap-free soft cloth, tray
1. Remove the contra-angle from the handpiece, place it in a clean
tray, and then remove the disposable isolation sleeve from the handpiece.
2. Soak the nap-free soft cloth with distilled water or deionized water,
and then wipe all the surfaces of the components such as the handpiece,
charger, base, etc. until the surface of the component is not stained.
3. Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces
of the handpiece, charger, base and other components for 3 minutes.
4. Put the handpiece, charger, base and other components back into
the clean storage area.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.
7.4 The cleaning, disinfection and sterilization of contra-angle,lip
hook, file clip, protective silicon cover, touch probeare as follow.
Unless otherwise stated, they will be hereinafter referred to as
“products”.
Warnings:
The use of strong detergent and disinfectant (alkaline pH>9 or acid
pH <5) will reduce the life span of products. And in such cases, the
manufacturer takes no responsibility.
The products may not be exposed to temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization
cycles. The materials used in manufacture were selected accordingly.
However with every renewed preparation for use, thermal and chemical
stresses will result in ageing of the products. The maximum number of
sterilizations for products is 250 times.
7.4.1 Initial processing
7.4.1.1 Processing principles
It is only possible to carry out effective sterilization after the
completion of effective cleaning and disinfection. Please ensure that, as
part of your responsibility for the sterility of products during use, only