Geophysical Survey Systems, Inc.
PaveScan® RDM 2.0
Manual
MN36-716 Rev A
20
The image below indicates that the calibration procedure is ready to begin with the airwave measurement.
1
Press the
Air
button. When pressed, it turns orange while the calibration measurement is being
performed. The air calibration takes 5-10 seconds per sensor.
•
For a 3-sensor set-up, the total measurement time is about 15-30 seconds.
•
All sensors must be in the air-measurement position during the calibration.
2
Once the air calibration is completed, the Metal Plate Calibration step can be initiated. One of the
buttons turns green. Make certain that the position indicated on the button matches the position of the
sensor on the cart. In the example below, sensor #12 should be on the left side of the cart. If it is not,
then the system should be shutdown and restarted with changes made to the Sensor Configuration
(2.2).
3
Place the metal plate underneath the sensor indicated. Be sure to line up the
metal plate so that the sensor laser is directed into the small hole in the
center of the metal plate.
The metal plate and cart also need to be on a
level surface or surface with a non-changing slope. If the metal plate is
not flush with the surface, move to another location where it is flush.
4
Once the metal plate is in place, lower the
antenna, and press the green button. It turns orange
during calibration. The metal plate calibration takes
from 5-10 seconds. Once it is complete, a different
button turns green.
The order in which
the buttons turn
green may vary.
Summary of Contents for PaveScan RDM 2.0
Page 1: ...MN36 716 Rev A ...
Page 4: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A ...
Page 6: ...Geophysical Survey Systems Inc PaveScan 2 0 RDM Manual ...
Page 8: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 2 ...
Page 34: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 28 ...
Page 46: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 40 ...
Page 50: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 44 ...
Page 52: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 46 ...
Page 64: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 58 ...
Page 66: ...Geophysical Survey Systems Inc PaveScan RDM 2 0 Manual MN36 716 Rev A 60 ...
Page 67: ......