IEC 61000-4-5
±2 kV common mode
±2 kV common mode
be that of a
typical
commercial,
hospital, or
residential
environment.
Voltage Dips, Short
Interruptions and Voltage
Variations on Power Supply
Lines
IEC 61000-4-11
<5% UT
(>95% dip in
U
T) for 0.5 cycle
40% UT
(60% dip in
U
T) for 5 cycles
70% UT
(30% dip in
U
T) for 25 cycles
5% UT
(>95% dip in
U
T) for 5 sec
<5% UT
(>95% dip in
U
T) for 0.5
cycle
40% UT
(60% dip in
U
T) for 5
cycles
70% UT
(30% dip in
U
T) for 25
cycles
5% UT
(>95% dip in
U
T) for 5
sec
Mains power
quality should
be that of a
typical
commercial,
hospital, or
residential
environment.
If the user of
the GSI Pello
requires
continued
operation
during power
mains
interruptions,
it is
recommended
that the Pello
be powered
from an
uninterrupted
power supply.
Power Frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power
frequency
magnetic
fields should
be at levels
characteristic
of a typical
location in a
typical
commercial
or hospital
environment.
Note: U
T is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The GSI Pello is intended for use in the electromagnetic environment specified below. The customer or the user of
the Pello should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Electromagnetic Environment-
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz to 80
MHz
3 V/m 80 MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Pello, including cables than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.