Pressure alarm delay
This pressure alarm delay allows for small pressure fluctuations without
trigging the high pressure alarm. When the WMS is sampling, the pressure
alarm delay sets the length of time that can expire before the high-pressure
alarm is triggered because of high pressure.
Default setting: 0 seconds
Range: 0–4 seconds
ATTENTION
Changing the high pressure warning/alarm/delay set-points can cause
system failure and will void the system warranty.
4.4 Alarm action
Some of the system alarms have the ability to stop the sampling process e.g.
stop pump or close motor valve if the alarms are activated. Figure 4.9 shows
the flow and pressure alarm settings. Moving the slider to "ON" will activate
flow and pressure alarms. To access these functions, press the three dots
located next to the alarm on the system settings page Figure 4.8.
Figure 4.9: Alarm action