Water Monitoring System
User Manual
Page
56
of 82
Ver. 1.1 – Revision March 2022
1. Perform PAH sensor check as described in section 7.4.3
2. Go to “Settings” menu. Login. Select “G6111 PAH”. Select “Validation”
Figure 7.12: G6111 PAH pop-up menu
Figure 7.13: PAH validation
Enter PAH liquid concentration value based on the sensor range, please
select the validation liquid in accordance with Table 7.1.
3. Take the syringe and attach the needle. Break the neck of the ampule.
Fill the syringe with the entire content of the ampule. Work quickly once
the ampule is opened as this will ensure the best results.