Water Monitoring System
User Manual
Page
22
of 82
Ver. 1.1 – Revision March 2022
When pressing the Calibration or Validation buttons the user will be
instructed in how to perform the selected function (refer to section 7).
4.3 System settings
The system settings screen can be accessed from the settings screen
(see Figure 4.6). Any functions that are not explained in Figure 4.1 will be
described in this section.
Figure 4.8: System settings
Remote start delay
Air might be present in the scrubber piping system at start-up. This
sampling delay ensures that air does not enter the system during the start-
up phase. The On/Off valve or the pump will turn on after the Remote Start
Delay period has expired. This function is
only
active in remote operation.
Remote Start Delay also ensures a degree of protection of the pump against
dry running. Therefore, changing the timer setting should be considered
carefully.
Default setting: 120 seconds
Range: 0–240 seconds
Flow establish time
The WMS starts sampling when the On/Off valve or the pump is activated.
Flow Establish Time allows for sufficient time for the water to travel from the
On/Off valve or the pump to the flow sensor. If the flow sensor is triggered
within the time set by the Flow Establish Time, the WMS will continue
sampling.