Validation & Calibration
Page
49
of 82
Figure 7.8: G6120 Turbidity pop-up menu
9. Follow the instruction on screen as shown in Figure 7.9 using standard
10 NTU or follow steps 10-12 below.
Figure 7.9: Turbidity validation
10. Enter the turbidity concentration used and press the “Start Validation”
button.
11. Wait for the display to update the validation result. If the validation
is successful, a calibration is not required. When the validation is
completed, go to step 30.