Page 111 of 112
Immunity test
IEC / EN 60601
test level
Compliance level
Electromagnetic
environment
–
guidance
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2700 MHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any parts of the GSI 39,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
P
d
2
,
1
=
P
d
2
,
1
=
80 MHz to 800 MHz
P
d
3
,
2
=
800 MHz to 2700 MHz
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, (a) should be less
than the compliance level in
each frequency range (b)
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
(a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the GSI 39 is used exceeds the applicable RF compliance level