To review the individual test results, press the
button and enter “
Page mode
.”
The word
“Page” will be displayed
in the center of the LCD. Testing cannot be performed while the system is in
the Page Mode. The memory number is located in the upper right-hand corner of each screen. If, for
example, only five tests were stored in memory, only five memory locations can be viewed. The
memory can be reviewed one page at a time by pressing the or button once
and observing the result. The entire memory can be scrolled through by holding the or
buttons down continuously. Press the
button to exit “Page Mode” and conti
nue testing.
Memory erase
If there is a particular test result that must be to delete before printing, enter the Page mode by
pressing . Press
or
to display the test result and press
. This erases that particular
test result from memory. The LCD displays a blank screen for erased memories with the memory
location number located at the top right corner. Upon exiting from the Page mode, the stored
memories reshuffle and replace the empty memory with the remaining tests in the order in which
they were obtained. The Page mode will be exited when you press the
PRINT ALL
or
ERASE
ALL
buttons or any button that would normally begin the setup of a new test. Page Mode is
ready only. No changes can be made to audiometric results.
To erase all tests from memory, press the
ERASE ALL
button.
Printing test results
The printout will begin with a header, if it is selected, in the program mode (i.e., GSI 39 or a custom
header). The next two lines contain space for recording the individuals name and the test date. This is
followed by the test results in the order they were obtained/ selected.
Either a single test can be printed from memory or the entire group of tests in memory can be printed.
To print a single test from memory, use the
PAGE
button to enter the Page mode and
the
or
button to arrive at the desired test result to print. When this test is displayed, press the
PRINT SCREEN
button.
To print all tests in memory, press the
PRINT ALL
button. When
PRINT ALL
is pressed
and two audiogram tests are stored in memory, they will combine under the following conditions.
There must be one left test and one right test sequentially stored in memory. A left and right
audiometric pair of tests will not be combined if they are separated in the memory by a Tymp test.
Therefore, when tests are erased, the result could cause a change in (left, right) or (right, left)
sequence with Audiometric tests. This would result in the wrong audiometric tests being combined
when
PRINT ALL
is selected. Prior to selecting
PRINT ALL
scroll through the tests
in memory to determine where the audiometric tests are located.
To avoid accidental confusion of data,
ERASE ALL
before starting a new test patient.