Page 102 of 112
4.
Contralateral is available with Versions 2 and 3 only.
Pressure:
226 Hz Probe Tone:
Reflex measures are set automatically to
pressure at peak compliance with an offset of -20 daPa if peak
pressure is negative and +20 daPa if peak pressure is positive.
1000 Hz Probe Tone:
Reflex measures are taken at 0 daPa regardless
of peak pressure.
Reflex Determination:
226 Hz:
Compliance change of 0.05 cm
3
or greater.
1000 Hz
: Compliance change of 0.1 mmho.
Reflex Test Time:
1 to 12 seconds depending upon the number of ipsilateral and/or
contralateral test frequencies selected (4 maximum) and intensity
required.
Probe LED Indicators
226 Hz Probe:
Steady yellow:
Occlusion
Blinking green:
Ready to start
Steady green:
Test in progress
Steady orange:
Leak in pressure
No Light:
Test is finished
Combo Probe:
Blinking green:
Ready to start
Steady green:
Test in progress
Steady orange:
Occlusion
Blinking orange:
Leak in pressure
No light:
Test is finished
Audiometry mode (Versions 3 and 4 only)
Frequencies:
125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000,
8000 Hz
Accuracy:
±2%
Total Harmonic Distortion:
< 2.5% (125 to 3000 Hz measured acoustically at maximum dB HL;
4000 and 6000 Hz measured electrically)
TRANSDUCERS
Audiometric Headset:
Pair DD45 earphones with Type 51 cushions
(10 ohms impedance)
Headband force per ANSI S3.6 and IEC 645 (4.5 ±0.5)
Insert Earphones:
ER-3A (10 ohm impedance)