Page 96 of 112
NO DATA AVAILABLE
This message appears if the
button is pressed and no results are stored.
NOT AVAILABLE
This message appears if the
button is pressed and the computer is not properly connected.
Data Transfer Program Mode
The Data Transfer Program mode is used to modify the GSI 39 USB interface configuration parameters
to match the computer
’
s USB Port settings.
Enter the Program mode by selecting the
PROG
ram button. Move the cursor to
Data Xfer Config
and
press the
button to enter the submenu.
The following screen appears the first time the Data Transfer Program mode is entered showing
the factory default settings.
PROGRAM MODE - DATA TRANSFER
* 115.2 kBAUD
* NO 8-BIT
57.6 kBAUD
ODD 7-BIT
38.4 kBAUD
EVN 7-BIT
19.2 kBAUD
SPC 7-BIT
9600 BAUD
* XON/XOFF DISABLED
4800 BAUD
XON/XOFF ENABLED
These selections fall into three groups:
•
Baud rate
•
Parity and data bits
•
Flow control
The default setting for each group has an asterisk (
*
) before it so that it is easy to scan the settings for
each group.
Selecting default settings for any of the groups is achieved in the same manner as the Program mode.
Use the or buttons to move the solid square cursor down or up to the setting that you wish
to select. Press the
button. The word
SAVED
will appear in the lower right corner of the screen
and the asterisk (*) will appear in front of the new setting.
To exit the Data Transfer Program submenu, move the cursor to the press the
button.
This will return to the Program mode which can be exited by selecting
PROG
ram.