- 42 -
7. Click the “X” button on the upper left on the Calibration window to close it,
and remove the calibration sheet from the CE7000-40 and finish the calibration.
8. After calibration is finished, perform the “Cut Test” to check the calibration of the F-Mar2 is correctly.
NOTICE
Two calibration sheets are included in the F-Mark2, if used up these, please print the
“caliblack.pdf” image data which is installed to the accessory CD to the label media,
and use it as a calibration sheet.
The calibration data is saved in the operation computer.
When change the operation computer, needs calibration with changed computer.
When USB cable connection was disconnected, needs calibration again.
When performing calibration under the strong lighting such as spotlight, calibration
may not be preformed correctly.