- 41 -
6. Click the “Read Maker” button to read the calibration data, and then click the “Set” button to saving
the calibration data.
NOTICE
When performing “Read Maker”.
If the “Read Marker” is performed while the calibration sheet is reflecting by the
Lighting in the room, the correct calibration may not be performed,
in this case, please shade the camera unit so that the small square part is not reflected
by the light, and then perform the “Read Maker”.
After performed Calibration.
When correct calibration is performed , the value of *90 to 120 is displayed in the
“Delta X” and the value of *95 to 135 is displayed in the “Delta Y”.
(*The values are ideal values, there varies depending on the equipment.)
If values for Delta X and Delta Y are much different than the above ideal values,
the calibration was not performed correctly. Please calibration again.