CONSERVATION AND CORRECTIVE AND PREVENTIVE
MAINTENANCE
Lubrication of the counter-angle
- Only lubricate the counter-angle (a), with a specific spray.
- Lubricate after each use and before the sterilization.
- Insert the tip (c) on the body of the atomizer.
- Insert the Spray adapter (b) on the rear part of the counter-angle (a) and lubricate
for 2 to 3 seconds or until lubricant comes out through the head of the counter-angle (a).
- Before inserting the lubricated counter-angle in the motorized handpiece, clean the
excess of oil with absorbent paper. Put it on an inverted or inclined position so that the
gravity helps to drain.
- Plug it after the excess of oil has been flowed out.
a
c
b
01
ATTENTION
- Never immerse the instrument in a disinfection bath.
- Never use microabrasive material or steel wool during cleaning.
These processes shall be used on handpieces.
- Uncouple the part from the fitting terminal.
- Remove the drill or diamond tip.
- Clean the external part using 70% alcohol or other disinfectant that does
not have acid, ammonia or chlorine.
Disinfection
Lubricant
For the lubrication of your
equipment, we recommend the use
of the Gnatus GTS product.
- The lubricant does not
accompany the product.
- The use of another brand
of lubricant will cause the loss of
warranty of the product.
WARNING
In order to avoid contamination, use gloves and other protection systems during
disinfection.
Even that, during the procedures, gloves are used, after their removal the hands
should be washed completely.
Summary of Contents for Endus Duo R
Page 1: ......
Page 12: ...12 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 13: ...13 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 14: ...14 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 15: ...15 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 66: ...66 ...
Page 67: ...67 ...
Page 68: ...68 ...