- The equipment should only be operated by duly enabled and trained technicians (Dental
Surgeons, Capacitated Professionals).
- If any maintenance should be required, only use services of the Gnatus Authorized
Technical Assistance.
- The equipment has been manufactured to handle both continuous.
- Although this equipment has been planned in accordance with the standards of
electromagnetic compatibility, it can, in very extreme conditions, cause interference with
other equipment. Do not use this equipment together with other devices very sensitive to
interference or with devices which create high electromagnetic disturbance.
- When the file accidentally touches the internal part of the canal of the dental root, the
length meter of canal advancement will become a bit abnormal, but, after a few seconds,
it will automatically go back to normal.
- Do not use damaged file supports; a precise measurement cannot be made with a
damaged file-holder.
- Only use files and file support with plastic cables. If the file has a metallic cable, there
will be a current leakage when the fingers touch the cable, which will impede a precise
measurement of the root canal. Even if the file cable is made of plastic, check to not touch
the metallic part with the fingers.
- Follow the manufacturer’s instructions for the use of the endodontics files.
- Do not use designated files for continuous spinning in reciprocating movement.
- Do not use designated files for reciprocating movement in continuous spinning.
- The precision of the movement proportioned by the motor is only assured if you use
an original counter-angle, properly maintained and lubricated.
- When lubricating the counter-angle, have special care so that the lubricant does not
penetrate on the motorized handpiece.
- Do not lubricate the motorized handpiece under any circumstances, since the contact
with the lubricant may damage it and have a strongly negative effect in terms of safety
during use.
- The motorized handpiece can heat up excessively if an excessive force is used. If the
motorized handpiece overheats with frequency or if the overheating persists, contact your
Gnatus representative.
Precautions and warnings “during use” of the equipment
- Position the equipment in a place where it will not get wet.
- Install the equipment in a place where it will not be damaged by pressure, temperature,
humidity, direct sunlight, dust, salts, or sulfur compounds.
- The equipment should not be submitted to inclination, excessive vibrations, or blows
(including during transportation and handling).
- This equipment was not planned for use in an environment where vapors, anesthetic
mixtures infl ammable with air, or oxygen and nitrous oxide can be detected.
- Before turning the equipment ON, check the electrical network. The transformer shall
be plugged in an outlet whose voltage is within the interval: 100 - 240 V 50/60 Hz.
Use original parts only.
- In general, the batteries fully charged gradually discharge with time even though the
device is not in use. It is recommended that the battery be charged immediately before use.
- Before its first use and/or long work interruptions such as vacations, clean and disinfect
the equipment (see Page 61 “Conservation and corrective and preventive maintenance” –
General cleaning, Disinfection and Lubrication).
Precautions and warnings “during the installation” of the
equipment
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Summary of Contents for Endus Duo R
Page 1: ......
Page 12: ...12 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 13: ...13 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 14: ...14 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 15: ...15 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 66: ...66 ...
Page 67: ...67 ...
Page 68: ...68 ...