54
M2:
With this function enabled, when pressing the
command pedal (a) or button (b) on the own
handpiece the motor will spin clockwise. If,
during its operation, the force reaches the pre-
defined limit concerning the torque, two beeps
will sound and the handpiece will automatically
spin in the reverse direction. Once the force is
eliminated, the handpiece starts to spin clockwise
again. Activating the REVERSE mode, beeps
will sound alerting the alteration in the spinning
direction, spinning it counterclockwise, when
disactivating and activating again the equipment
the spinning direction will go back to the initial
clockwise direction (NORMAL).
a
b
M 2
AUTO REV
Function (M2) – Inversion of the Automatic Spinning
2.7
N.cm
Load within the torque limit value
More load beyond the configured
torque limit value
If the applied load is maintained,
the file will spin inversely (reverse).
When the load is removed
Forward spinning
Reverse spinning
Forward spinning
After the 0,5mm scale the beeps shall be short and frequent reaching the ideal level
“APEX”, where this word will be blinking on the panel, when the tip of the instrument reaches
the foramen.
ATTENTION: If the instrument outreaches the apex, the word “OVER” (which represents
the exceeding area of the apex) will be blinking followed by intermittent beeps.
The labial clip can be disconnected from the instrument and connected again during the
measurement cycle. It can be exchanged for one or another more adequate during the
root canal treatment or when another canal is measured. In these cases the equipment
automatically detects that a new measurement has been initiated.
EQUIPMENT OPERATION
NOTE
- The length of work on the canal varies from case to case due to different forms of the
teeth and canals of the dental roots;
- After the length of work has been defined turn off the equipment and disconnect all
cables.
Summary of Contents for Endus Duo R
Page 1: ......
Page 12: ...12 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 13: ...13 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 14: ...14 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 15: ...15 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Page 66: ...66 ...
Page 67: ...67 ...
Page 68: ...68 ...