GlideScope
®
Cobalt AVL System
User’s Manual
37
Cleaning the Cobalt AVL
Table 3: Chemical Compatibility and Disinfection
Methods for Cobalt AVL Video Baton
Enzymatic
debridement
agent/
detergent
General
hospital grade
As per chemical
manufacturer's
instructions
N/A
Surface
cleaning only
in preparation
for disinfectant
Isopropyl
Alcohol
Solution
70%
70% used to
wipe down with
minimum 1 minute
exposure
Low
Glutaraldehyde
Up to 3.4%
2.0% – exposure
for minimum
20 minutes at
20 °C or as per
manufacturer’s
instructions
High
Ortho-
Phthalaldehyde
0.55%
0.55% – exposure
for 12 minutes
at 20 °C or as per
manufacturer’s
instructions
High
Peracetic Acid
0.2%
0.2% – exposure
for minimum
12 minutes at 5`
56 °C or as per
manufacturer’s
instructions
See
comments
on right
Classified as
a chemical
sterilant
Bleach
(Sodium
Hypochlorite)
Up to
8000 ppm
5000 ppm –
exposure for 10
minutes at 20 °C
High
Corrosive for
connector pins
and SS ring
500 ppm used to
wipe down with
minimum 1 minute
exposure
Low
Noncorrosive
at ≤ 500 ppm
Active
Ingredient
Compatibility
Conditions
Disinfection
Level
Caution/
Comments
Refer to page 41 for Disinfecting Steps.
Hydrogen
Peroxide
Up to 7.5%
7.5% – exposure
for 30 minutes at
≥20 ºC or as per
manufacturer's
instructions
High
Summary of Contents for Cobalt AVL
Page 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Page 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Page 26: ......
Page 27: ...GlideScope Cobalt AVL System User s Manual...
Page 28: ......
Page 29: ...GlideScope Cobalt AVL System User s Manual...
Page 30: ......
Page 77: ......