Replacing the GlideScope Monitor Battery
Under normal operating conditions, the monitor battery will last 2–3 years;
or approximately 500 charge/discharge cycles.
The battery is not user-replaceable. In case of battery malfunction, do not
attempt to replace the monitor battery. Any attempts to replace the battery
by unauthorized service technicians may cause serious harm to the user
and will void the warranty. Please contact your Verathon Medical Customer
Care Representative for more information on battery replacement.
Device Disposal
Disposal of this device can be coordinated through your Verathon Medical
Service Center in accordance with WEEE requirements.
Cleaning the GlideScope
Cobalt AVL Monitor
Wipe the exterior of the monitor with IPA (70% isopropyl alcohol), bleach
(100 ppm) or a mild detergent and water.
Cleaning the GlideScope
Cobalt AVL Cradle
Wipe the cradle with a standard hospital grade surface cleaning product.
Disinfecting the GlideScope
Cobalt AVL Video
Baton
To disinfect use any of the chemicals listed in Table 2 and do the following:
1. Ensure the equipment is clean (see Cleaning Section for video baton).
2. Ensure the protective caps on the connectors are secure.
3. Prepare the disinfectant solution at the concentration and temperature
recommended by the disinfectant manufacturer. Disinfect the
equipment following the disinfectant manufacturer’s instructions or as
stated in Table 1.
4. After the disinfection process, rinse (as applicable) and dry the video
baton and then store the equipment in a clean environment.
Summary of Contents for Cobalt AVL
Page 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Page 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Page 26: ......
Page 27: ...GlideScope Cobalt AVL System User s Manual...
Page 28: ......
Page 29: ...GlideScope Cobalt AVL System User s Manual...
Page 30: ......
Page 77: ......