GlideScope
®
Cobalt AVL System
38
User’s Manual
Cleaning the Cobalt AVL
Table 4: Chemical Compatibility and Disinfection
Methods for GlideRite
®
Rigid Stylet
Active
Ingredient
Compatibility
Conditions
Disinfection
Level
Caution/
Comments
Glutaraldehyde
Up to 3.4%
2.0% – exposure
for minimum
20 minutes at
20 °C or as per
manufacturer’s
instructions
High
Ortho-
Phthalaldehyde
0.55%
0.55% –
exposure for
12 minutes at
20 °C or as per
manufacturer’s
instructions
High
Peracetic Acid
0.2%
0.2% – exposure
for minimum 12
minutes at 50 to
56 °C or as per
manufacturer’s
instructions
See
comments
on right
Classified as
a chemical
sterilant
Isopropyl
Alcohol
Solution
70%
70% – exposure
for 10 minutes at
20 °C
Intermediate
70% used to
wipe down with
minimum 1
minute exposure
Low
Bleach
(Sodium
Hypochlorite)
≤ 500 ppm
500 ppm used
to wipe down
with minimum 1
minute exposure
Low
Non-
corrosive at
≤ 500 ppm
Enzymatic
debridement
agent/
detergent
General
hospital
grade
As per chemical
manufacturer's
instructions
N/A
Surface
cleaning only in
preparation for
disinfectant
N/A
Tested in
autoclave
(steam cycle)
Minimum
4 minute
132 °C pre-
vacuum steam
sterilization
cycle
See
comments
on right
Based on requests
from users,
an autoclave
cycle has been
established for
the Stylet for ease
of use
Refer to page 42 for Cleaning and Disinfecting Steps.
Summary of Contents for Cobalt AVL
Page 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Page 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Page 26: ......
Page 27: ...GlideScope Cobalt AVL System User s Manual...
Page 28: ......
Page 29: ...GlideScope Cobalt AVL System User s Manual...
Page 30: ......
Page 77: ......