GlideScope
®
Cobalt AVL System
36
User’s Manual
General Cleaning and Disinfection Information
Table 2: Risk Assessment
Cleaning the Cobalt AVL
Warning:
Disinfectants and cleaning methods listed are recommended by Verathon
®
based on compatibility with product materials. Refer to the disinfectant
label instructions for guidance on disinfection efficacy and appropriate
clinical uses.
Caution:
Meticulous cleaning must precede any disinfection process, to
ensure all foreign matter is removed from the surface of the device. This
allows the active ingredients of the chosen process to reach all the surfaces
of the device.
Availability of disinfection products varies by country, and we are unable
to test products in every market. Please use the list of recommended
disinfectants in this manual to compare with products available locally.
Note:
When using any of the high-level disinfectants listed in Table 3, read
and comply with product use instructions in all applications.
Device
Sterile
Use
Spaulding’s/
CDC
Classification
Disinfection
Sterilization
Low
Level
Int.
Level
High
Level
Video
Baton*
Non-
sterile
Reusable
Noncritical
X
GVL
®
Stat
Sterile
Single-
use
Semicritical
Stylet
Non-
sterile
Reusable
Semicritical
X
Monitor
#
Non-
sterile
Reusable
Noncritical
X
Cradle
#
Non-
sterile
Reusable
Noncritical
X
It is understood that all items in this chart will be used as intended.
Note:
* The Cobalt AVL Video Baton is a nonsterile, reusable device, which is protected from contact
with mucous membranes and non intact skin by the Stat (sterile, single-use) when used as
intended. Low Level Disinfection is recommended for the Cobalt AVL Video Baton after every
patient use. High Level Disinfection is required for the video baton when it is visibly soiled.
# Disinfect monitor and cradle when they are visibly soiled and on a regular basis as per a
schedule established by the medical care facility or provider.
– Shaded areas—not required/not compatible with device materials.
– Checked boxes (X) show minimum requirement.
– Unshaded areas show permissible levels of disinfection based on compatibility with device
materials.
Summary of Contents for Cobalt AVL
Page 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Page 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Page 26: ......
Page 27: ...GlideScope Cobalt AVL System User s Manual...
Page 28: ......
Page 29: ...GlideScope Cobalt AVL System User s Manual...
Page 30: ......
Page 77: ......