8
1. Screen: Displays waves, menu, alarm and all measuring parameters.
2.
/ : Function button:
a) When menu (except the TREND menu) is activated, press this
button to move the cursor.
b) When the TREND menu is activated, this button changes between
the trend graph and data table
On the main display, to press this button to silence alarms
for 2 minutes.
3.
: Press this button to move the cursor when menu is activated.
4.
+
: Multifunction button.
a) Press this button to increase figures on the menu.
b) In the main display screen, to press this button to freeze the display
wavefor m (if frozen, the data to print is the data of the frozen moment)
5.
-
: Press this button to decrease figures.
6.
ENTER
: Confirmation button;
a) Press this button to “Confirm” on the menu.
b) In the main menu, to press this button to turn on the pump directly
if it has automatically time switched OFF
c) If the device is connecting with Bluetooth printer, press this button
for 2 seconds to print capnography and other result parameters
(EtCO
2
, RR, SpO
2
, PR).
7.
Press this button to enter or quit menu or change display
8. Power button: hold for >2 seconds to activate
9. Indicator
POWER
: Blue LED is lit when the Monitor is either switched
ON or subject to external power when not switched ON.
If the yellow LED is lit, the internal battery is being charged.
10.
CO
2
: The faucet of filter, blue color indicator flashes if the filter T3 is
off. when the filter T3 is plugged in ,the indicator color will change to be
blue all the time, and it will change to be red during occlusion or pump
err.
11.
SpO
2
: The socket of SpO
2
(optional).
12. DC5V Mini USB Charging interface.
Note
: this interface must only be connected to a device which meets
safety standards.
13. Exhaust outlet: Do not occlude.
14. Speaker location
15. Battery Compartment with clip on Battery Door
16. Hanging Point for Lanyard if required.