16
III. CO
2
SET Menu
Figure 7
In this menu, press
or
button to move the cursor up or down, press +
button or - button to change the data highlighted by the cursor.
To return to the main menu highlight EXIT and press the
ENTER
button. If you want to return the monitor to its default settings highlight
LOAD DEFAULTS and Press the
ENTER
button.
This menu includes the following setups:
1. The high alarm limits of
EtCO
2
: EtCO
2
ALARM_H
: 22-99mmHg, off
2. The low alarm limits of
EtCO
2
: EtCO
2
ALARM_L
: off, 10-60mmHg
3. The high alarm limits of respiration rate:
RESP ALARM_H
:
5-60 breaths/min, off
4. The low alarm limits of respiration rate:
RESP ALARM_L
: off,
4-40 breaths/min
5. Pump flow rate setup:
FLOW-SET
: 50cc/min-250cc/min
6. The setup of apnea time:
APNEA TIME
: 15s-44s, off
7. The unit of CO
2
:
CO
2
UNIT
: %, mmHg or kPa
8. Pump switch:
ON
or
OFF
9. Pump auto-closing time:
AUTO-OFF-TIME
: 10-30min
10. Screen speed of capnograph:
SWEEP SPEED
: SLOW, NORMAL or FAST
11. CO
2
Wave scale:
WAVE SCALE
: 54mmHG or 76mmHG
12. EtCO
2
average computation time:
EtCO
2
Averaging
: every breath,
10sec, 20sec, 30sec
13. Default reload:
LOAD-DEFAULTS
14. Exit:
EXIT
Attention:
Pump auto-closing time means that the pump will automatically be closed
down when no respiration occurs in the set period (default 10 min).