24
Please do not let alcohol, cleaning reagent or sterilizing reagent into
filter/water trap. Check that the filter/water trap’s internal membrane is dry
clean before it is used. Although recommended as a single patient use
item for maximized infection control, in an emergency the filter/water trap
may be re-used if in good condition. In this case the filter/water trap
should be removed, the surface should be cleaned as in section III below,
with a dry tissue and the filter/water trap then placed in ventilated dry
environment to let it dry naturally. Replace the filter/water trap if it is dirty,
shows any sign of contamination or if in any doubt about its condition.
Please Note: It is advised to use the filter/water trap T3 as a single
patient use item so as to absolutely remove the risk of crossinfection.
Attention:
The sample line or water trap should be not sterilized and used repeatedly
if the packing indication shows that it is disposable.
Attention
:
For the environment, disposable sample line or filter shall be treated suitably
or recycled.
III. Cleaning
Warning: Before to clean the monitor and probe, power shall be turned
off and removed from any charging source.
1. Cleaning the Monitor
It is recommended that the Monitor is used in the supplied Carry Case
which offers protection from both contamination, liquid ingress and damage.
Do Not to sterilize the Monitor by high pressure, autoclave or washer.
Do Not to dip or expose the Monitor to liquid.
Do Not to use the Monitor if there is any sign of damage.
Use only PH Neutral Cleaning products.
This product is not suitable for mechanical re-processing or sterilization.
Monitor Cleaning Instructions: Only the Carry Case and if necessary
the Monitor surfaces may be cleaned and/or disinfected. Use Moist
(not dripping) wipes with 70% solution of isopropyl alcohol, or very dilute
Chlor-clean (1000ppm) or Chlorohexidine (1000ppm) or mild detergent,
then allow to air dry naturally.
2. Cleaning the SpO
2
probe.
Care:
Do Not sterilize the probe by high pressure, autoclave or washer.
Do Not dip the probe into liquid.
Do Not use the probe if there is any sign of damage.
Use only PH Neutral Cleaning products.