26
6
7
8
Attention: Please to contact the Distributors Customer Service if
you require advice, replacement parts and/ or service.
Appendix 1.
Explanations of Terms in this Manual
MENU
Menu
EtCO
2
The CO
2
concentration of expiration end phase
INCO
2
The CO
2
concentration of inspiration phase
SpO
2
Oxygen saturation
RR or R-RATE
Respiration rate
PR or P-RATE
Pulse rate
mmHg
Millimeters Mercury
kPa
Kilopascal
ALARM-H
Alarm high limit
ALARM-L
Alarm low limit
LINE
Line curve
FILL
Filled or solid under waveform
BEEP_VOLUME
Pulse volume
ALARM_VOLUME
Alarm volume
LOW BATTERY
Low battery charge
APNEA
Apnea or breathing stopped for a set period of time
Breaths/Min
Breaths per minute
CAL
Offset Calibration
SET
Setup
2. Screen indicating TEMP-
LOW.
3. Screen indicating
TEMP-IMBALANCE.
No values of SpO
2
or no wave.
Flashing ‘LOW BATTERY’ and
closed down automatically.
Still flashing ‘LOW BATTERY’
after the power is supplied
and AC indicator no light.
1. Finger too cold.
2. Interference of very strong
external light.
3. The measurement test
of SpO
2
and blood pressure
are done on the same arm.
4. Red light in the sensor
no flashing.
5. Infrared and collector
of sensor is not clean.
1. No Battery Charge.
1. Battery Charger power
working abnormally.
1. Warm up finger.
2. Avoid strong external
light.
3. Place SpO
2
sensor on
other arm or position.
4. Renew SpO
2
sensor.
5. Clean internal parts of
SpO
2
Sensor.
1. Connect to Battery
Charger.
1. Check battery charger
and cable and replace as
necessary.