10
Figure 3
Instruction for use of the filter/water trap T3:
1. Insert the convex cleat of filter/water trap T3 into the notch of the inlet port
of device and turn T3 45°clockwise.
2. Attach male luer lock sample line connector to filter/water trap T3
(Use a Male to Male Luer adapter if the sample line has Female Luer
connector)
3. Connect the other end of the Sample line to the chosen sampling point
of Patient or Ventilator Circuit.
4. Change the filter/water trap T3 as needed – This item may dried if very wet
and re-used, subject to local Infection Control requirements. If the
filter/water trap T3 becomes dirty or the Occlusion Alarm is activated
when it is dry then the filter/water trap must be replaced.
WARNING
Ensure that connections are air tight as if there is leakage, measured values
are likely to be inaccurate.
WARNING
Use only recommended Original bespoke filter/water trap T3 to ensure
accuracy.
II. Respiration rate measurement
The calculation of respiration rate derives from monitoring the CO
2
waveform.