Installation of Small Clip
Follow the step 1, step 3 and step 4 mentioned in Section Installation of Big Clip and refer to Figure 7A
and Figure 7B.
Figure 7A Figure 7B
Small Clip Installation
Installation of Rubber Wrapper
1. Spread the rubber wrapper, there are two fixing slots (Figure 8A);
2. Hold the rubber wrapper and press one branch of the Y-type sensor into one of the fixing slots (Figure
8B), then fix the other branch.
Figure 8A Figure 8B
Rubber Wrapper Installation
3.3 Placement of Veterinary SpO2 Sensor
As a general rule, measuring SpO2 and pulse rate for animal is performed when the animal is calm or
under anaesthesia, and the first choice of measuring site is their ears, then the second choice is their
tongues. Normally the animal's tongue will slip out after anaesthesia. At first, clip the sensor on its ear
(Figure 9) or tongue (Figure 10). Then, according to perfusion index (PI) value (greater PI is better) you
can find an appropriate site.
Besides the ear and tongue, you can wrap the SpO2 sensor onto other places such as leg (Figure 11) or tail
(Figure 12). Since animal hair will decrease the sensitivity or even make the measurement failure, please
always choose the place with less hair under the sensor. If necessary, shave off the hair on the measuring site.
Figure 9 clip the sensor on tongue Figure 10 clip the sensor on tongue