26
ENGLISH
At this time select “Yes” and press “OK” key to delete all the selected records.
5 Technical Specifications
A. Display panel:
color LCD;
B. Power supply:
3x LR6 (AA) alkaline batteries or Ni-MH rechargeable batteries
Supply voltage: 3.2~5.0VDC
Operating current: <150mA
Continuous working time: >30 hours
C. SpO2 Specifications
Transducer: dual-wavelength LED sensor
Measurement wavelength:
Red light: 660 nm, Infrared light: 905 nm.
Maximal optical output power: less than 2mW maximum average
Measuring range: 35~100%
Measuring accuracy: Not greater than 3% for SpO2 range from 70% to 100%
*NOTE: Accuracy defined as root-mean-square value of deviation according to ISO 9919.
Low alert setting range: 50%~99% (Default setting: 80%).
D. Pulse Rate Specifications
Range: 30bpm~400bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
Alert setting range: 25bpm~400bpm
Default alert setting: High: 140bpm Low: 50bpm
E. Perfusion Index Display
Range: 0~20%
F. Update rate
8 beats moving average for both SpO2 and pulse rate readings
G. Data Recording
Store data every 1/2/4/8 seconds, up to 384–hour data record.
H. Low perfusion performance
The accuracy of SpO2 and PR measurement still meet the precision described above when the modulation
amplitude is as low as 0.5%.
I. Dimensions:
145 mm (L) × 74 mm (W) × 29 mm (H)
Net Weight: 210g (including batteries)
J. Classification
The type of protection against electric shock: Internally powered equipment
The degree of protection against electric shock: Type BF applied parts.
The degree of protection against harmful ingress of liquids: Ordinary equipment without protection against
ingress of water.
Electro-Magnetic Compatibility: Group I, Class B
6 Accessories
1. A set of veterinary probe
2. A holder
3. Battery (AA) × 3
4. A User Manual
5. A Quality Certificate
6. A data cable (optional)
7. Oximeter Data Manager software (optional)
Note: The accessories are subject to change. See the Packing List for detailed items and
quantity.