4.4.1 Setup
On the menu screen, select “
SETUP
” and then press “
” key for entering into system setup screen.
The setup screen is as shown in the following figures.
Figure 17 System Setup Screen (A)
Figure 17 System Setup Screen (B)
Operation Instructions:
1.
DATE
: Date setting
1) When cursor stays on the Year of the date, press “
” (Mode/OK) key to active Year option, the
cursor flashes on the Year of the date;
2) Press
/
(Navigation key) to adjust year.
3) Press “
” (Power/back) key or “
” (Mode/OK) key to confirm and exit from date setting.
4) The procedures of adjusting Month value and Day value are the same with Year adjustment.
Date Format: yy-mm-dd
Note: The setting operations of other parameters (such as TIME, RECORDING INTERVAL,
POWER SAVING etc.) are the same with date setting.
2.
TIME
: Time setting
3.
Thin Tissue
: Setting measurement option for thin tissue (without bone) such as ear pr tongue; two
options:"YES" and "NO". The factory default setting is "NO", which is for thick tissue option such as leg
or tail.
4.
Alarm-SpO2
: SpO2 alarm setting; the factory default value is 80%.
5.
Alarm-PR Hi
: Pulse Rate High limit setting, the factory default value is 140.
6.
Alarm-PR Lo
: Pulse Rate Low limit setting, the factory default value is 50.
7.
RECORDING INTERVAL
: Record interval setting; five options: “1s, 2s, 4s, 8s” and “OFF”; When the
option is “OFF”, the device will not record real-time measurement data.
Note: The length of data record is limited to 30 seconds at least, and the maximal length
for one record is also limited to one hour (for one second interval) , 2hours (for 2 seconds
interval) 4 hours (for 4 seconds interval) or 8 hours (for 8 seconds interval).
8.
POWER SAVING
: power saving setting; two options: "on" and "off". The factory default setting is "on".