Figure 11 wrap the sensor on leg Figure 12 wrap the sensor on tail
Tips for SpO2 Monitoring
A. If there is strong ambient light interference, you can use opaque thing to cover the measuring site and
the sensor. By dong so, measurement error may be decreased.
B. The light including surgery lamp, halogen lamp, fluorescent lamp and infrared heating lamp can also
cause measurement error. In such condition, you can use solid black thing to cover the measuring site
and the sensor.
C. For thin tissue (without bone) such as tongue or ear, please make sure the measurement option "Thin
Tissue" in the setup menu is selected as "YES"
4 Operation
4.1 Default Screen
Press “
” power key for 2 seconds to start the Oximeter, then insert finger into rubber cushions of the
probe, the screen will display the default screen, as shown in Figure 13.
Figure 13 Default Screen
Screen Description:
“%SpO2” “99”
: SpO2 value; “99” is a percentage value;
“19:18”
: Current time;
“PR 65”
: Pulse rate value: 65bpm;
“
”
: Battery voltage indication;
“ ”
: Data memory icon; when the data on screen becomes stable, the device will start to store data
automatically. No icon means the current data will not be stored. If the memory is full or the total number
of the records reaches 256 pieces, the earliest record will be overwritten and the icon “ ” will appear
on the screen for prompt, as shown in the figure below.
Note: It is suggested that the data shall be uploaded to computer for saving, or the earliest
record will be overwritten.
Memory full icon