ENGLISH
ENGLISH
ENGLISH
ENGLISH
S
UPER
D
EE
-
SV
20
3.
OPERATIVE MENU
3.1
LANGUAGE SELECTION MENU
To select a language different than the one currently set, operate as follows:
Switch off the distillator with the ON/OFF SWITCH.
Switch on again pressing simultaneously the
SHIFT
key on the keyboard and keep pressed until
the end of the initial test.
The display shows the current language and the shifting arrows in correspondence of
A
and
C
keys. Select the language with these two keys.
Confirm with the
MODE
key
.
3.2
OPERATIVE MENU
See also Chapter 5 (ANALYSIS – Example of distillation)
The instrument display is composed by two lines of text. The first shows the operative mode and the
second describes the function corresponding to
A
,
B
and
C
keys.
Switching on the instrument, the display shows the standard setpoints of the distillate to collect (see
Figure 3).
By pressing the
MODE
key, it is possible to select the quantity:
Set : 80
80 160 240
The software has 3 definite setpoints :
80
,
160
and
240
grams, selectable by pressing the
A
,
B
or
C
key
respectively
The precision balance shows the gram.
Set : 87
- +
It is possible to modify these values by pressing the
SHIFT
key with the
A
(-) or
B
(+) key
By pressing the
MODE
key, it is possible to proceed to the distillation start menu,
Standby
Start
It is possible to start the distillation by pressing the A key,
correspondent to the START function. The distillation does not
start without a volumetric flask or a conical flask on the balance.
Once the distillation has started, the display shows the weight value collected and the total volume to be
collected.
15 -> 80
Stop
When the distillate reaches the selected quantity (80), the
instrument does not accept a new start if the flask is not
removed.
It is possible, at any moment, to stop the distillation by pressing the
B
key (
Stop
). If the distillation is
running, only the STOP key (
B
) is active.