2610 Calibration
Calibration
settings:
1
Click to Start Dissolved Oxygen Saturation
2
Click to Reset Dissolved Oxygen Calibration to Factory Default
First Point
settings:
3
With probe in 100% Saturated Water click Start to begin the Calibration
4
Once probe reading is stable click Continue
Second Point
settings:
5
With probe in 0% Saturated Water click Start to begin the second point Calibration
6
Once probe reading is stable click Finish to complete calibration
Reset Calibration:
7
Click Reset Cal to reset the DO calibration to Factory Default
1
2
3
4
5
6
7
15
GF 0252 Configuration Tool